Nov 24, 2021 10:29
2 yrs ago
14 viewers *
English term

concordance specification

English to Russian Medical Medical: Instruments
ORI Accuracy shall meet a concordance specification of a minimum of 80% with directional changes in PaO2 over a range of 100-200 mmHg.

Discussion

Alexander Kondorsky (asker) Nov 24, 2021:
Пульсоксиметр Masimo
Tymur Buliekov Nov 24, 2021:
Александр, а что за прибор?

Proposed translations

1 hr
Selected

*

точность измерения ORi должна отвечать требованиям соответствия направления изменения ORi направлению изменения PaO2 на уровне не хуже 80%
https://techdocs.masimo.com/globalassets/techdocs/pdf/lab-92...
Directional ORi changes
greater than 0.04 have been validated to greater than 80% concordance to directional
changes in PaO2 greater than 10mmHg.
Something went wrong...
4 KudoZ points awarded for this answer. Comment: "Спасибо!"
14 mins

технические требования по соответствию

-
Something went wrong...